Dongfang Jinghuayuan SGH-300/500/1000 High Purity Hydrogen Generator Hydrogen Gas Chromatography Laboratory

USD 599.90

Dongfang Jinghuayuan SGH-300/500/1000 High Purity Hydrogen Generator Hydrogen Gas Chromatography Laboratory

Description

Note: The machine is heavier, and the logistics is self-pickup

1. Overview and working principle: This instrument is mainly composed of electrolysis system, pressure control system, purification system and display system. Electrolyzed hydrogen adopts the current membrane separation technology, a pressure control system composed of an infrared photoelectric feedback device and a switching power supply, which can automatically adjust the flow of hydrogen according to the needs of the output and maintain the stability of the output pressure. 1. The appearance of the instrument:

 

2,工作原理图:

 

3. Instrument features: (1) It can replace the high-pressure hydrogen cylinder to make the laboratory instrumentation and ensure safety. (2) Fully automatic control of the working process, simple operation and convenient daily maintenance. (3) The digital display of hydrogen production is convenient for observing the working status and fault judgment of the instrument. (4) Long service life, continuous or intermittent use, stable gas production, no attenuation. (5) It is equipped with over-voltage protection device, which is safe and reliable. 2. Main parameters: 1. Output flow: -300 0-300ml/min; 2. Output pressure: 0.4Mpa (factory setting) 3. Hydrogen purity: >99.999% 4. Maximum power:-300 150W 5. Working conditions: (1) Power supply: AC 220V ±10%; 50Hz (2) Ambient temperature: 5-40℃ Relative humidity: <85% No serious dust pollution in the environment 6. Dimensions: 390×185×320 (L×W×H) 7. Weight: 15kg 3. How to use: After taking out the instrument, check the appearance and interior of the instrument for any abnormal phenomena such as serious damage caused by transportation. Preparation before starting: (1) To ensure the normal use of this instrument, the operator needs to prepare the following items: 1. 2 liters of distilled water 2. A funnel 3. A 500ml beaker 4. A glass rod 5. AC220V power socket (flat three-wire; 10A) to ensure reliable grounding. 6. Stainless steel pipe or copper pipe with outer diameter φ3mm. (2) Prepare electrolyte: Dilute 100 grams of analytical pure potassium hydroxide (KOH) with 400-500ml of distilled water, dissolve and cool, and pour it into the liquid tank, and then add distilled water to the "1.2" mark (not exceeding). Note: When adding electrolyte, use a funnel to slowly inject it to avoid outflow. (3) Tighten the sealing cap of the hydrogen output port on the rear panel of the instrument. (4) Check the grounding end of the power cord of the instrument to make it grounded reliably. Power on: (1) Turn on the power, turn on the switch, and observe the flow display. At this time, the displayed value should be: -300 type 320-400ml/min

(2) When the pressure gradually rises and the pressure reaches 0.4Mpa (set value), the digital display returns to zero and the display is "000-020", indicating that the instrument is normal. Online: After the startup is normal, turn off the machine, then remove the sealing pressure cap on the hydrogen outlet, connect the gas chromatograph with a pipe with an outer diameter of φ3mm, and ensure that the sealing joint does not leak, then start the instrument, and wait until the pressure reaches the set pressure be usable. Every time you use it in the future, you only need to turn on the power switch of the instrument. Fourth, the replacement of color-changing silica gel: After using the instrument for a period of time, it is found that the silica gel in the drying tube gradually changes color from bottom to top. When it exceeds a certain height, the silica gel should be replaced. The operation steps are as follows: Rotate in the direction shown, Unscrew the cover of the drying tube:

 

 

2. Take out the inner tube of the drying tube upwards, Take out the cotton ball and pour out the silica gel.

 

3. Install new (or dried) silica gel, the amount should not exceed the small vent near the nozzle. Cover the cotton ball, be careful not to have cotton thread and other sundries on the end of the drying tube (otherwise it will cause air leakage), and then put it into the outer tube of the drying tube. 4. Tighten the cover of the drying tube, and use soap to check for leaks after starting the machine to ensure the seal. Five, matters needing attention: The instrument should be placed on the workbench for easy observation and operation. This instrument is not allowed to run for a long time (above 10min) when the pressure is zero, otherwise it will cause abnormal electrolyte circulation. Observe the discoloration of the desiccant frequently, and replace the discoloration silica gel in time. Observe the liquid level frequently and add distilled water in time to keep the liquid level between 0.6-1.2 scale lines. Since this instrument only consumes distilled water when working, it does not consume KOH, so just add distilled water. (It is recommended to replace the electrolyte once a year. For the preparation method of the new electrolyte, refer to "3, 3, (2),"). If the electrolyte is lower than the 0.6 mark, it will cause the electrolytic cell to burn. The hydrogen flow value displayed by this instrument is for reference only. 6. Packing list: One host. Analyze 100 grams of pure potassium hydroxide (KOH). One ear-washing ball; one meter of liquid-pumping plastic tube. A copy of the instruction manual; Six rubber sealing rings φ6×1.9, one φ45×3.5; Two M8×1 sealing caps; two insurance tubes (2A);

 

Attachment: Simple faults and troubleshooting methods:

 

 

序号

Failure phenomenon

Judgement of the cause of failure

Solution

1

The flow rate is displayed as the maximum flow rate, and the pressure does not rise

Block the output port with a sealing pressure cap, turn it on and observe whether it can reach the set pressure and return to zero.

Can return to zero. Then the external air circuit leaks.

Check the joints of the external air circuit with soap liquid, and seal the leaks.

Do not return to zero. then Air leakage inside the instrument.

Use soap to check both ends of the drying tube and the internal joints of the instrument, and seal the leak.

2

After reaching the set pressure, the flow display is very different from the actual use flow;

3

No hydrogen is produced after starting up, and the digital display does not light up.

The instrument fuse is burned out. The electrolytic power supply fuse is burned out.
  1. .

Replace the fuse

4

No hydrogen is produced after starting up, the digital display is "000"

  1. The electrolysis power supply stops vibrating. The light lotus root table is in poor contact. The pressure protection switch is protected.
  1. 更换电源或压Replace the power supply or pressure switch. Plug the optical lotus connection
 

 

Specification

Type : Hand Tool Parts

Origin : Mainland China

+